We fix these points with the customer before a test. The examples come from tests we have run.
- Cell chemistry
- NCM, LFP, LCO. Pouch cells and prismatic cells.
- Sample
- Modules and packs. Examples: 3P4S LFP pack, 710.4 Wh LCO pack, 1128.5 Wh NCM module.
- Trigger
- Overcharge (electrical abuse), a heating device between cells (5–10 °C/min), or both together.
- Agent application
- Fixed water-mist system (4 MPa for 240 s) or portable extinguisher (6 L), released when the first cell goes into thermal runaway.
- Measurement
- Thermocouples across the pack (12 channels in the fixed-system test), video of the whole run, cell voltages after the observation time, agent consumption.
- Comparison
- The same procedure repeated with deionized water and with another agent, so that the curves can be compared.
- Acceptance criteria
- Written before the test. Examples: flame out, no re-ignition within 20 minutes or 1 hour, at least one cell group still at 3.7 V or more, agent consumption within the limit.
- Reference procedures
- KFI criteria No. 630 for small lithium-ion battery fires, NTA 8133+A1, and the Chinese technical requirements for fire control devices of electric-bus battery boxes.